Reflectivity and Transmittance of Extremely low loss optical components GLACIER

This articles discusses the reflectivity and transmittance measurements of extremely low loss optical components (reflectivity 99.9995%) through the use of a CRD Loss Meter and Reflectometer.


Cavity-Ringdown (CRD) Loss Meter and Reflectometer

Our reflectometer GLACIER uses the extreme sensitivity of cavity ring-down spectroscopy to 

quantify the losses of advanced optical coatings down to 5 ppm. As a typical application the 

device can characterize supra-mirrors with up to 99.9995 % reflectivity. Conventional 

absorption and reflection measurements are not sufficiently sensitive to quantify today´s 

super-reflective mirror coatings and are typically limited to the > 1000 ppm range 

(corresponding to < 99.9 % reflectivity). Cavity ring-down spectroscopy measures optical

losses by the decay of the energy stored inside a cavity. The technique reaches unrivalled 

sensitivity, because losses are experienced with each round trip inside the cavity over and over 

again. Smaller losses lead to longer intra-cavity dwell time thereby automatically increasing 

measurement precision. The device features high-speed data acquisition and allows to record 

measurements within seconds. It is delivered complete with a computer and a user-friendly 

software interface for acquisition and real-time analysis.


image.png


GLACIER

Key Product Features

  Reflectivity measurements up to 99.9995%

  Various angles of incidence: 5°-45° (and 0°)

  S and p polarization (separately)

  Antireflective coating characterization down to 0.0005% (5 ppm)

  Simple and reproducible alignment for 0.5”, 1” and 2” optics

  Spring-loaded mirror fixtures for reproducible mounting without strain

  Computer and user-friendly software interface included

  High-speed data acquisition and real-time analysis

  Wavelengths in the range 375-1550 nm available

  Footprint 90 x 45 cm²